All pipettes have been quality tested according to ISO8655-2:2002 with calibration certificate. The quality control involves gravimetric testing of each pipette with distilled water at 22ºC.
Features for MicroPette
- Lightweight, ergonomic, low force design
- Digital display clearly reads volume setting
- The pipettes cover volume range of 0.1μl to 10ml
- Easy to calibrate and maintain with tool supplied
- Design helps avoid repetitive strain injuries
- Calibrated in accordance with ISO8655. Each pipette supplied with individual test certificate
- The low part is available for autoclaving
- Manufactured from innovative materials
Features for multi-channel MicroPette
- 8 and 12 channel pipettes are available for standard 96-well plate
- Dispensing head rotates for optimum pipetting convenience
- Individual piston and tip cone assemblies allow easy repair and maintenance
- Compound material tip cone design allows visual seal verification
- Can be used with universal style pipette tips
 |  |
Calibration Please visit our website for online calibration at www.dragon-lab.com. Through the online calibration software, users can perform simple, accurate and timely calibration with free cost, and avoid the calibration error due to non-professional operation. Â We will support and help you to achieve consistently excellent results. |
 |
Specifications
Â
 Â
Single-channel Adjustable Volume Pipettes | |||||||||||
Volume Range | Increment | Test Volume | Maximum permissible systematic error (Inaccuracy) | Maximum permissible random error (Imprecision) | |||||||
% | μL | % | μL | ||||||||
0.1-2.5μL | 0.05μL | 2.5μL | 2.50% | 0.0625 | 2.00% | 0.05 | |||||
1.25μL | 3.00% | 0.0375 | 3.00% | 0.0375 | |||||||
0.25μL | 12.00% | 0.03 | 6.00% | 0.015 | |||||||
0.5-10μL | 0.1μL | 10μL | 1.00% | 0.1 | 0.80% | 0.08 | |||||
5μL | 1.50% | 0.075 | 1.50% | 0.075 | |||||||
1μL | 2.50% | 0.025 | 1.50% | 0.015 | |||||||
2-20μL | 0.5μL | 20μL | 0.90% | 0.18 | 0.40% | 0.08 | |||||
10μL | 1.20% | 0.12 | 1.00% | 0.1 | |||||||
2μL | 3.00% | 0.06 | 2.00% | 0.04 | |||||||
5-50μL | 0.5μL | 50μL | 0.60% | 0.3 | 0.30% | 0.15 | |||||
25μL | 0.90% | 0.225 | 0.60% | 0.15 | |||||||
5μL | 2.00% | 0.1 | 2.00% | 0.1 | |||||||
10-100μL | 1μL | 100μL | 0.80% | 0.8 | 0.15% | 0.15 | |||||
50μL | 1.00% | 0.5 | 0.40% | 0.2 | |||||||
10μL | 3.00% | 0.3 | 1.50% | 0.15 | |||||||
20-200μL | 1μL | 200μL | 0.60% | 1.2 | 0.15% | 0.3 | |||||
100μL | 0.80% | 0.8 | 0.30% | 0.3 | |||||||
20μL | 3.00% | 0.6 | 1.00% | 0.2 | |||||||
50-200μL | 1μL | 200μL | 0.60% | 1.2 | 0.15% | 0.3 | |||||
100μL | 0.80% | 0.8 | 0.30% | 0.3 | |||||||
50μL | 1.00% | 0.5 | 0.40% | 0.2 | |||||||
100-1000μL | 5μL | 1000μL | 0.60% | 6 | 0.20% | 2 | |||||
500μL | 0.70% | 3.5 | 0.25% | 1.25 | |||||||
100μL | 2.00% | 2 | 0.70% | 0.7 | |||||||
200-1000μL | 5μL | 1000μL | 0.60% | 6 | 0.20% | 2 | |||||
500μL | 0.70% | 3.5 | 0.25% | 1.25 | |||||||
200μL | 0.90% | 1.8 | 0.30% | 0.6 | |||||||
1000-5000μL | 50μL | 5000μL | 0.50% | 25 | 0.15% | 7.5 | |||||
2500μL | 0.60% | 15 | 0.30% | 7.5 | |||||||
1000μL | 0.70% | 7 | 0.30% | 3 | |||||||
2-10mL | 0.1mL | 10mL | 0.60% | 60 | 0.20% | 20 | |||||
5mL | 1.20% | 60 | 0.30% | 15 | |||||||
2mL | 3.00% | 60 | 0.60% | 12 | |||||||
8-channel Adjustable Volume Pipettes | |||||||||||
Volume Range | Increment | Test Volume | Maximum permissible systematic error (Inaccuracy) | Maximum permissible random error (Imprecision) | |||||||
% | μL | % | μL | ||||||||
0.5-10μL | 0.1μL | 10μL | 1.50% | 0.15 | 1.50% | 0.15 | |||||
5μL | 2.50% | 0.125 | 2.50% | 0.125 | |||||||
1μL | 4.00% | 0.04 | 4.00% | 0.04 | |||||||
5-50μL | 0.5μL | 50μL | 1.00% | 0.5 | 0.50% | 0.25 | |||||
25μL | 1.50% | 0.375 | 1.00% | 0.25 | |||||||
5μL | 3.00% | 0.15 | 2.00% | 0.1 | |||||||
50-300μL | 5μL | 300μL | 0.70% | 2.1 | 0.25% | 0.75 | |||||
150μL | 1.00% | 1.5 | 0.50% | 0.75 | |||||||
50μL | 1.50% | 0.75 | 0.80% | 0.4 | |||||||
12-channel Adjustable Volume Pipettes | |||||||||||
Volume Range | Increment | Test Volume | Maximum permissible systematic error (Inaccuracy) | Maximum permissible random error (Imprecision) | |||||||
% | μL | % | μL | ||||||||
0.5-10μL | 0.1μL | 10μL | 1.50% | 0.15 | 1.50% | 0.15 | |||||
5μL | 2.50% | 0.125 | 2.50% | 0.125 | |||||||
1μL | 4.00% | 0.04 | 4.00% | 0.04 | |||||||
5-50μL | 0.5μL | 50μL | 1.00% | 0.5 | 0.50% | 0.25 | |||||
25μL | 1.50% | 0.375 | 1.00% | 0.25 | |||||||
5μL | 3.00% | 0.15 | 2.00% | 0.1 | |||||||
50-300μL | 5μL | 300μL | 0.70% | 2.1 | 0.25% | 0.75 | |||||
150μL | 1.00% | 1.5 | 0.50% | 0.75 | |||||||
50μL | 1.50% | 0.75 | 0.80% | 0.4 | |||||||
Fixed Volume Pipettes | |||||||||||
Volume Range | Increment | Test Volume | Maximum permissible systematic error (Inaccuracy) | Maximum permissible random error (Imprecision) | |||||||
% | μL | % | μL | ||||||||
5μL | - | 5μL | 1.3% | 0.065 | 1.2% | 0.06 | |||||
10μL | - | 10μL | 0.8% | 0.08 | 0.8% | 0.08 | |||||
20μL | - | 20μL | 0.6% | 0.12 | 0.5% | 0.1 | |||||
25μL | - | 25μL | 0.5% | 0.125 | 0.3% | 0.075 | |||||
50μL | - | 50μL | 0.5% | 0.25 | 0.3% | 0.15 | |||||
100μL | - | 100μL | 0.5% | 0.5 | 0.3% | 0.3 | |||||
200μL | - | 200μL | 0.4% | 0.8 | 0.2% | 0.4 | |||||
250μL | -- | 250μL | 0.4% | 1.0 | 0.2% | 0.5 | |||||
500μL | - | 500μL | 0.3% | 1.5 | 0.2% | 1.0 | |||||
1000μL | - | 1000μL | 0.3% | 3.0 | 0.2% | 2.0 | |||||
2000μL | - | 2000μL | 0.3% | 6.0 | 0.15% | 3.0 | |||||
5000μL | - | 5000μL | 0.3% | 15 | 0.15% | 7.5 |
Â
Massey Ferguson is a manufacturer of agricultural equipment, formed by the 1953 merger of farm machinery manufacturers Massey Harris of Canada and the Ferguson Company in Britain.And now Holdwell Group also supply the replacement parts for Massey Ferguson Engine.Holdwell Group stock many Engine Parts in the warehouse and gives you very fast delivery time,and also we can develop engine parts for you,ensure you get the parts in cheap price but also in good quality,You will have wonderful experience in Holdwell Group.