Automatic calorimeter on the installation and use of oxygenation meter

There are many accessories in the calorimeter, the main ones being the oxygenator and the oxygen bomb. In this chapter, the technicians of our coal testing equipment will briefly introduce the installation and use of the oxygenator. I hope the new and old customers will have a general understanding. Skilled operation:

1. Before installation, carefully inspect each part for tightness, whether there is any trace of damage and damage to the appearance, and all loose parts should be intact.

2. Connect the oxygenator to the oxygen relief valve with an oxygenating catheter and tighten all nuts. Do not loosen.

3. Open the oxygen cylinder main valve and adjust the pressure reducing valve so that the pressure gauge is displayed as 2.8-3.0 MPa. At this time, there should be no air leakage phenomenon in the entire air path, otherwise it should be checked until it is normal.

4. The oxygenation test should be conducted in the world. At this time, it should be free from leakage and operate freely. The pressure gauge on the oxygenator should be consistent with the pressure on the decompression valve.

5, open the oxygen tank total valve (bottle pressure should be greater than 5MPa), the oxygen bomb placed on the oxygenator positioning seat, the oxygen warhead aligned oxygenator gas nozzle, down oxygen pressure handle, in 2.8-3.0 Oxygen under pressure MPa for 30 to 40s and then slowly fly, release the handle to remove the oxygen bomb.

After the oxygenation instrument is installed correctly, the oxygen bomb can be used normally, and the test can be completed. Therefore, the oxygenation instrument installation is very important to determine whether the test can be completed and whether it can be tested accurately. This means that new and old customers must first learn to install Calorimeter oxygenator.